Inverted Metallurgical Microscope
ECLIPSE MA200
Instructions
M498E 08.10.NF.1 (2/4)
Zobrazit stránku 0
1 2 3 4 5 6 ... 106 107

Shrnutí obsahu

Strany 1 - ECLIPSE MA200

Inverted Metallurgical Microscope ECLIPSE MA200 Instructions M498E 08.10.NF.1 (2/4)

Strany 2

Contents 8 3.13.2 Adjusting the polarizer direction ...

Strany 3 - Introduction

5Troubleshooting 98 Improper use of the microscope may adversely affect performance, even if the microscope is not damaged. If any of the problems

Strany 4 - Safety Precautions

Chapter 5 Troubleshooting 99 Problem Cause Countermeasure Dirt or dust is seen in the view-field. The aperture diaphragm is stopped down too far.

Strany 5 - WARNING

Chapter 5 Troubleshooting 100 Problem Cause Countermeasure The lamp voltage is too low. Adjust the brightness with the brightness control dial.

Strany 6 - 8. Air vents

Chapter 5 Troubleshooting 101 Problem Cause Countermeasure A wrong objective is used. Place an appropriate objective into the optical path. (p.3

Strany 7 - CAUTION

6Care and Maintenance 102 Nikon recommends daily care and maintenance for maintaining the performance as long as possible. Do not let dust, fingerp

Strany 8 - 8. Cable routing

Chapter 6 Care and Maintenance 103 6.3 Storage • Store this product in a dry place where mold is not likely to form. • Store the objectives and

Strany 9

7Specifications 104 Type ECLIPSE MA200 Dimension 413 (W) × 337 (D) × 308 (H) mm, including the handle and mount parts Weight 20 kg Optical system

Strany 10 - Contents

Chapter 7 Specifications 105 Power cord When used in 100-120 V region, outside Japan: UL listed detachable power cord set, 3 conductor grounding (3

Strany 11

Contents 9 4.3 Network Connection...

Strany 12 - Part Name and Function

1Part Name and Function 10 Figure 1.1-1 Front view, Right-side view Eyepiece tube (Figure illustrates the MA2-TI3.) Binocular eye

Strany 13

Chapter 1 Part Name and Function 11 Figure 1.1-2 Front view Display POWER LED Displays ON/OFF for the power switch or the light-status

Strany 14

Chapter 1 Part Name and Function 12 Figure 1.1-3 Left view, Rear view Binocular tube Vertical tube Back port Supporting Pilla

Strany 15

Chapter 1 Part Name and Function 13 Figure 1.1-4 Connectors on the rear About product parts names in this manual The right colu

Strany 16 - Microscopy

2Microscopy 14 This chapter describes the procedure of each microscopy. See Table 2.1 for the items required for each microscopy. • Refer to “4. A

Strany 17

Chapter 2 Microscopy 15 Table 2.1 Items required for the microscopy Microscopy Page Illuminator Revolving nosepiece Objective Other items req

Strany 18 - Chapter 2 Microscopy

Chapter 2 Microscopy 16 2.1 Bright-field Microscopy under the Episcopic Illumination Turn on the power to light up the episcopic illumination la

Strany 19

Chapter 2 Microscopy 17 Set the microscope for the bright-field microscopy under the episcopic illumination. If the current unit is the one shown i

Strany 21

Chapter 2 Microscopy 18 Set the desired magnification and observe the sample. 1 Turn the revolving nosepiece to place the objective of a desired m

Strany 22

Chapter 2 Microscopy 19 2.2 Dark-field Microscopy under the Episcopic Illumination Attach the items required for the dark-field microscopy under

Strany 23 - ■ Sensitive color microscopy

Chapter 2 Microscopy 20 2.3 Polarization Microscopy under the Episcopic Illumination (simplified/sensitive color) Attach the items required for t

Strany 24 - 2.5 Epi-fl Microscopy

Chapter 2 Microscopy 21 2.4 Differential Interference Contrast Microscopy under the Episcopic Illumination Attach the items required for the diff

Strany 25

Chapter 2 Microscopy 22 Return to the bright-field microscopy under the episcopic illumination. 1 Pull the polarizer/analyzer unit to the first

Strany 26 - (See Table 2.1.)

Chapter 2 Microscopy 23 Return to the bright-field or dark-field microscopy under the episcopic illumination. 1 Turn the BD field changeover lev

Strany 27

Chapter 2 Microscopy 24 2.6 Bright-field Microscopy under the Diascopic Illumination Attach the items required for the bright-field microscopy u

Strany 28 - Adjust the focus again

Chapter 2 Microscopy 25 Set the microscope for the bright-field microscopy under the diascopic illumination. When the “Other items required” (See t

Strany 29

Chapter 2 Microscopy 26 Place the sample onto the stage to adjust the focus. 1 Use the sample holder according to the selected sample. Place th

Strany 30

Chapter 2 Microscopy 27 Center the condenser. (See 3.23.2.) 1 Be sure that the objective is set to the 10X objective. If not, turn the revolv

Strany 31 - Operation Details

1 Introduction Thank you for purchasing a Nikon product. This instruction manual is written for users of Nikon Inverted Metallurgical Microscope E

Strany 32 - 3.2 Illumination

Chapter 2 Microscopy 28 2.7 Polarization Microscopy under the Diascopic Illumination (simplified/sensitive color) Attach the items required for t

Strany 33 - Chapter 3 Operation Details

3Operation Details 29 3.1 Power ON/OFF 3.1.1 Power of the microscope Figure 3.1-1 Power of the microscope The power switch for the product

Strany 34

Chapter 3 Operation Details 30 3.2 Illumination 3.2.1 Brightness control and illumination ON/OFF Figure 3.2-1 Brightness control and illumina

Strany 35 - 3.4 Eyepiece Tube

Chapter 3 Operation Details 31 3.2.3 Displaying the POWER LED Figure 3.2-3 Displaying the POWER LED The status of the illumination lamp app

Strany 36 - Eyelevel riser

Chapter 3 Operation Details 32 Table 3.3-1 Selecting the microscopy method Microscopy Revolving nosepiece Objective Items required for the opti

Strany 37 - 3.6 Adjusting the Diopters

Chapter 3 Operation Details 33 3.4 Eyepiece Tube 3.4.1 Selecting optical path • The light distribution for both the binocular and vertical tube

Strany 38

Chapter 3 Operation Details 34 3.4.2 Adjusting the eyelevel risers Eyelevel risers can be used to adjust the height of the eyepiece tube for the us

Strany 39 - ■ Sample holder

Chapter 3 Operation Details 35 3.6 Adjusting the Diopters Adjust the diopter according to the user's eyesight using the diopter adjustment r

Strany 40

Chapter 3 Operation Details 36 3.7 Adjusting the Focus (for focus operation) 3.7.1 Using the coarse/fine focus knob Focus on the sample lifting/

Strany 41

Chapter 3 Operation Details 37 3.8 Placing the Sample and Operating the Stage 3.8.1 Placing the sample Figure 3.8-1 Placing the sample Nor

Strany 42 - 3.10 Filter

2 Safety Precautions To ensure correct and safe operation, read this manual before using the product. Warning and Caution Symbols Used in This Manua

Strany 43

Chapter 3 Operation Details 38 3.8.2 Changing the Observation Position The rectangular stage is equipped with the knob to move the stage in all fo

Strany 44

Chapter 3 Operation Details 39 3.9 Operating the Revolving Nosepiece and the Objective 3.9.1 Revolving nosepiece in combination with the objecti

Strany 45

Chapter 3 Operation Details 40 3.10 Filter Figure 3.10-1 Filter turret Two quartet filter turrets are provided on the microscope, and each filt

Strany 46

Chapter 3 Operation Details 41 3.11 Adjusting the Field Diaphragm (for the episcopic Illumination) Figure 3.11-1 Field Diaphragm Adjustment

Strany 47

Chapter 3 Operation Details 42 3.12 Adjusting the Aperture Diaphragm (for the episcopic illumination) Figure 3.12-1 Aperture diaphragm adj

Strany 48

Chapter 3 Operation Details 43 Adjustment with centering telescope (C-CT) Remove one of the eyepieces and attach the centering telescope instead.

Strany 49 - 3.15.3 Interference color

Chapter 3 Operation Details 44 3.13.2 Adjusting the polarizer direction Each time you turn the polarizer rotation ring, the direction of the polar

Strany 50 - 3.16.3 Interference color

Chapter 3 Operation Details 45 3.14 Using the λ Plate (MA2-λP) Figure 3.14-1 λ plate slider Attach the λ plate to the polarizer/analyzer u

Strany 51

Chapter 3 Operation Details 46 3.15 Using the DIC Slider (L-DIHC/L-DIC) Figure 3.15-1 L-DIHC/L-DIC slider Attach the L-DIHC/L-DIC DIC sl

Strany 52

Chapter 3 Operation Details 47 3.15.3 Interference color You can change the interference color continuously by turning the prism movement knob. Inte

Strany 53

Safety Precautions 3 WARNING 1. Intended product use The product should only be used for microscopic observation. Do not use this microscope for o

Strany 54

Chapter 3 Operation Details 48 3.16.3 Interference color You can change the interference color continuously by adjusting the orientation of the pol

Strany 55

Chapter 3 Operation Details 49 3.18 Using the λ Plate (D-LP) Figure 3.18-1 λ plate slider Attach the D-LP λ plate to the sextuple revolvin

Strany 56 - Figure 3.20-2 Scale slider

Chapter 3 Operation Details 50 3.19 Using the Fluorescent Unit (MA2-FL) Attach the fluorescent unit to the MA200 main body to perform the epi-fl

Strany 57

Chapter 3 Operation Details 51 3.19.2 Excitation light filter (EX filter) An excitation light filter transmits lights selectively and blocks oth

Strany 58

Chapter 3 Operation Details 52 3.19.3 Barrier filter (BA filter) A barrier filter transmits only fluorescent lights emitted by the sample but bloc

Strany 59

Chapter 3 Operation Details 53 3.20 Using the Scale Slider (MA2-GR/MA2-MR) Attach the MA2-GR/MA2-MR slider to the MA200 main body to measure stee

Strany 60

Chapter 3 Operation Details 54 3.20.2 Scale slider (MA2-MR) Figure 3.20-2 Scale slider Three items of reticles including “magnification,” “length

Strany 61

Chapter 3 Operation Details 55 3.21 Using the Intermediate Magnification Unit (MA2-MC) Figure 3.21-1 Intermediate magnification unit Attach

Strany 62

Chapter 3 Operation Details 56 3.22 Using the HG Precentered Fiber Illuminator (C-HGFI/C-HGFIE) Figure 3.22-1 HG Precentered Fiber Illuminat

Strany 63

Chapter 3 Operation Details 57 3.23 Using the supporting pillar for dia-Illuminator 100W (MA2-DP) Attach the supporting pillar for dia-Illuminato

Strany 64

Safety Precautions 4 7. Heat from the light source The lamp and the lamphouse become extremely hot. To avoid burns, do not touch the lamphouse whil

Strany 65

Chapter 3 Operation Details 58 3.23.2 Focusing and centering the condenser To perform the diascopic illumination for the first time or to replace t

Strany 66

Chapter 3 Operation Details 59 3.23.3 Adjusting the field diaphragm (for the diascopic illumination) The field diaphragm is used to limit the irrad

Strany 67

Chapter 3 Operation Details 60 3.23.4 Adjusting the aperture diaphragm (for the diascopic illumination) Adjust the size of the aperture diaphra

Strany 68 - WD value of the objective

Chapter 3 Operation Details 61 Filter list Filter Description ND2 (Neutral Density filter) Adjusts the brightness for general microscopy or photomi

Strany 69

Chapter 3 Operation Details 62 3.23.8 Polarizer slider (T-P2), λ plate (TI-DIC) When performing the simplified polarization microscopy under the di

Strany 70 - Assembly

Chapter 3 Operation Details 63 Dark cross that appears in the objective pupil Figure 3.23-10 Adjusting the polarizer direction 7 Loosen the c

Strany 71 - Chapter 4 Assembly

Chapter 3 Operation Details 64 3.24 Using the Supporting Arm (MA2-MP) The supporting arm is designed to mount Nikon camera control unit DS-L2. Th

Strany 72 - 4.1 About the System

Chapter 3 Operation Details 65 3.25 Using the DS Camera Control Unit (DS-L2) Once you connect the DS Camera Control Unit DS-L2 with the DS camera

Strany 73

Chapter 3 Operation Details 66 Figure 3.25-2 Objective info Displaying the objective info Press the button, and the name and other info are dis

Strany 74

Chapter 3 Operation Details 67 3.25.5 Connecting with the DS camera head switcher (DS-SW: Camera switch BOX, optional) If the DS-L2 is connected wi

Strany 75 - [Note 1]

Safety Precautions 5 CAUTION 1. Handle with care This product is a precision optical instrument. Handle the microscope system with care to avoid sh

Strany 76

4Assembly 68 WARNING • Before assembling the microscope, be sure to read the WARNING and CAUTION at the beginning of this instruction ma

Strany 77

Chapter 4 Assembly 69 Combination of the illuminator and the light source Use this product with a specified light source. The specified light so

Strany 78

Chapter 4 Assembly 70 4.1 About the System D-LP λ plate slider D-DA analyzer slider Eyepiece D-ND6 revolving nosepiece D-NI7 revolving nosepi

Strany 79 - CAUTION

Chapter 4 Assembly 71 4.2 Combination List for the Unit Combinations for each microscopy provided for MA200. : All selectable or available. c: S

Strany 80 - 4.5 Revolving Nosepiece

Chapter 4 Assembly 72 1: For EPI objective on 5 revo., “LU nosepiece adapter M32-25” is required. 2: BD objective is unavailable on 6 revo., 7 rev

Strany 81

Chapter 4 Assembly 73 4.3 Network Connection 4.3.1 Unit system capable of establishing network The table below shows a unit system capable of es

Strany 82 - 4.6 Stage

Chapter 4 Assembly 74 4.3.2 Viewing user control options and status displays on a PC or the DS-L2 The table below shows user control options and s

Strany 83 - 4.9 Eyepieces

Chapter 4 Assembly 75 Table 4.3-1 User control options for the unit using the PC/DS-L2 PC DS-L2 Unit Control Status Control Status Brightness

Strany 84 - 4.11 Filters

Chapter 4 Assembly 76 4.3.3 Registering information on the objective You can register the information to be checked on a PC or the DS-L2 with the “

Strany 85

Chapter 4 Assembly 77 4.4 The LV-LH50PC Lamphouse and the Lamp CAUTION for heat Do not touch the lamp and the lamphouse while the lamp is o

Strany 86

Safety Precautions 6 • Do not hold the focus knobs, eyepiece tube, lamphouse, stage, and so on, when carrying the microscope. They may come off and

Strany 87 - 4.13.1 MA2-λP λ plate

Chapter 4 Assembly 78 4.4.2 Replacing the lamp 1 Loosen the lamphouse cover clamp screw using the supplied hexagonal wrench. 2 Remove the lamph

Strany 88

Chapter 4 Assembly 79 4.5.2 Cable connection for the manual revolving nosepiece and the MA200 Connect the cable of the revolving nosepiece to the c

Strany 89 - WARNING

Chapter 4 Assembly 80 4.6 Stage • To attach or remove the stage, hold the stage securely to prevent from being fallen. • To remove the stage f

Strany 90

Chapter 4 Assembly 81 4.7 Objectives 1 Remove the sample holder from the stage. 2 Screw the objectives into the socket of the revolving nosepi

Strany 91 - Preparation for attachment

Chapter 4 Assembly 82 4.10 Grain Scale Slider/Scale Slider 1 Remove the dust protection cover from the scale slider slot on the bottom-front of t

Strany 92

Chapter 4 Assembly 83 4.12 Various Sliders for Attaching the Revolving Nosepiece To insert a slider, be sure to attach the objective in advance.

Strany 93

Chapter 4 Assembly 84 4.12.2 LV-DIC/LV-DIHC slider (Senarmont method) Attach the DIC slider in Senarmont method to the D-ND6 revolving nosepiece.

Strany 94

Chapter 4 Assembly 85 4.13 Polarizer/analyzer Unit and Fluorescent Unit 1 Loosen the front-cover fixing bolt with the supplied hexagonal screwdr

Strany 95

Chapter 4 Assembly 86 4.14 Intermediate Magnification Unit Opening for the intermediate magnification unit 1 Remove four bolts to take out th

Strany 96

Chapter 4 Assembly 87 4.15 External Light Source You can attach the HG precentered fiber illuminator (for the episcopic illumination) or the supp

Strany 97 - ■ Attaching the DS-L2

7 Contents Introduction ...

Strany 98 - 4.17 Eyelevel Risers

Chapter 4 Assembly 88 * For the C-HGFIE motorized fiber illuminator, connect the C-HGFIE and the [HGFIE] connector on the MA200 using the RS232C cr

Strany 99 - 4.19 Power Cord

Chapter 4 Assembly 89 Attach the condenser to the condenser mount. You can attach the TI-C condenser turret (system condenser) or ELWD-S condenser

Strany 100 - Troubleshooting

Chapter 4 Assembly 90 Attaching the ELWD-S condenser With the indication on the turret facing the front (towards the user), insert the conden

Strany 101

Chapter 4 Assembly 91 Attach/replace the lamp. Be sure to use the halogen lamp (OSRAM HLX64623 or PHILIPS77241) exclusive for the D-LH/LC Lamphouse

Strany 102

Chapter 4 Assembly 92 Attach the filter and the slider. • Do not touch filters or other optical components with your bare hands. • Latches at bo

Strany 103

Chapter 4 Assembly 93 Attach the T-P2 polarizer slider. 1 Remove the system condenser, if attached, from the condenser holder to attach the

Strany 104 - Care and Maintenance

Chapter 4 Assembly 94 4.16 Supporting Arm (for DS-L2) You can mount the camera control unit DS-L2 on the MA200 main body by attaching the support

Strany 105 - 6.3 Storage

Chapter 4 Assembly 95 Attaching the DS-L2 1 Remove the one-touch release attached to the front of the mount arm for the supporting arm. Pull

Strany 106 - Specifications

Chapter 4 Assembly 96 4.17 Eyelevel Risers 1 Fully loosen the eyepiece tube clamp screw on the microscope, to mount the eyelevel riser onto the

Strany 107 - Chapter 7 Specifications

Chapter 4 Assembly 97 4.18 Camera Adapter You can perform photomicroscopy by mounting the photomicrographic equipment such as CCD camera on the m